Health ProductsType I recall

AK 98, 115V BIO/ AK 98, 115V SELF-CARE and AK98 V2 115V BIO VERSION / AK 98 V2 115V BIO SELF CARE/ AK 98 V3 (2018-08-29)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Gastroenterology and urology
Product: AK 98, 115V BIO/ AK 98, 115V SELF-CARE and AK98 V2 115V BIO VERSION / AK 98 V2 115V BIO SELF CARE/ AK 98 V3 (2018-08-29)
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Health professionals
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
All lots.955606AK 98 (SW 3)
All lots.115249
115251
AK 98
All lots.955406
955407
AK 98 (SW 2)

Issue

Baxter is communicating important safety information regarding the potential for AK 98 hemodialysis devices to generate excessive Ultrafiltration (UF) in certain situations where treatment-related alarms occur, or where there is an ultrafilter leak. Excessive UF may present hazards for sensitive patients, such as low-weight patients, for whom target UF values of zero or very low volumes are desired.