Health Products

API 50CH

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: API 50CH
Issue: Medical devices - Inaccurate test or measurement results
What to do: Immediately stop using the product and contact manufacturer.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
1008603700API 50CHN/A

Issue

During the internal investigation of the non-conformance opened on API 50 CH sequence 40-49 lots #1008352920 and 1008510240 for false positive results on test 49 it was confirmed that test 19 of API 50 CH lot# 1008603700 contains a mix of sorbitol(SOR) and 5-Ketogluconic acid (5kg, which is contained in the solution of the test 49) instead of sorbitol only. The mix is estimated to represent almost 80% SOR and 20% 5kg. During the investigation it was identified that test well 19 could be impacted on two API 50 CH lots, 1008603700 and 1008511380 and product stop shipment(PSS) was issued for both lots on 22 June 2021. The NMR analysis was performed on well 19 of both API 50 CH lots, this testing confirmed that lot 1008511380 is conforming.

Recall date: 2021-08-13

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime