Health Products

Protein C 30 & TOXO IGG Avidity 30

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices
Product: Protein C 30 & TOXO IGG Avidity 30
Issue: Medical devices - Performance
What to do: Immediately stop using the product and contact manufacturer.
Audience: General public, Health professionals, Hospitals
Recall class: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
1008571770
1008662180
1008801740
PROTEIN C 3030115
1008721750TOXO IGG AVIDITY 3030 222

Issue

Based on an unusual rate of complaints from the field for "Substrate error" leading to a potential delayed results on different References tested on all Systems, Biomerieux initiated an investigation. The substrate error issue was confirmed on all lots of Immuno-Assays manufactured with substrate batches using the common lot of raw material (4-MUP). The only potential risk associated with the substrate error is a delayed result as you need to run another test. There is no risk of false results. The investigation demonstrates that the issue occurs over the product shelf-life. The analyzes of the kinetic model allowed to define a revised expiry date for each impacted lot of Immuno-Assays finished products. When used until the revised expiry date, the product continues to perform per its registered performance specifications. Therefore, corrective actions involving a revised expiration date for all lots of clinical Immunoassay products are required to ensure that the specified products will continue to perform per their labelled performance specifications.

Recall date: 2021-10-19

Get told about the next health product recall.