Health ProductsType I recall

Surgical Microscope Leica M220 F12

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices
Product: Surgical Microscope Leica M220 F12
Issue: Medical devices - Mechanical defect
What to do: Immediately stop using the product and contact manufacturer.
Audience: General public, Health professionals, Hospitals
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
70414002
240414001
260311010
Surgical Microscope Leica M220 F1210448367

Issue

leica microsystems received one complaint (May 2021) pertaining to the M220 F12 surgical microscope system stating that the optics carrier fell down during the surgical procedure. The issue was found to be related to a loose screw in the swing arm of the M220 F12 which is holding the optics carrier in place. Based upon this knowledge and a previous recall for the same issue this recall was opened.

Recall start date: 2021-10-21