Health Products

Freelite Human Lambda Free Kit (for use on the SPAPLUS)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices
Product: Freelite Human Lambda Free Kit (for use on the SPAPLUS)
Issue: Medical devices - Inaccurate test or measurement results
What to do: Immediately stop using the product and contact manufacturer.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
Lot: 435251
Lot: 472776
FREELITE HUMAN LAMBDA FREE KIT,USE ON THE SPA PLUSLK 018.S
LK018.L.S

Issue

Following an investigation of customer complaints, an increase in curve activity and decrease in top point separation between calibrator (CAL) points 5 and 6 was observed for IK018.S lot 472776 (lK018.L.S lot 435251 shares the same components). Cal. 5 OD values may occasionally exceed Cal.6 Values, which may lead to imprecision within this region of the curve. Therefore, TBS will initiate the correction by way of issuing an UFSN to the affected customers.

Recall date: 2021-10-14

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime