ALLURA XPER FD10 (PRODUCT OF ALLURACLARITY FAMILY)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: ALLURA XPER FD10 (PRODUCT OF ALLURACLARITY FAMILY)
Issue: Medical devices - Software defect
What to do: The actions that should be taken by the user in order to prevent risks for patients or users •Ensure that the wired foot switch is always connected to the system. The wired footswitch is always provided with the AlluraXper and the Azurion systems. •Immediately start using the wired foot switch in case the wireless foot switch loses connection. •Circulate this notice to all users of this device so that they are aware of the issue. •Place this Urgent Medical Device Recall Letter with the documentation of the Philips AlluraXper or Azurion system. •Return the attached Urgent Medical Device Recall Letter Response Form to Philips to confirm that: o the Allura Xper and Azurion system wired foot switch is connected to the system o the users of the system have reviewed and understood this Urgent Medical Device Recall Letter.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
Not applicable.AZURION 7 B12722 067
Not applicable.AZURION 7 B20722 068
Not applicable.AZURION 7 M12722 078
722 223
Not applicable.AZURION 7 M20722 079
722 224
Not applicable.ALLURA XPER FD20 (PRODUCT OF ALLURACLARITY FAMILY)722028
Not applicable.ALLURA XPER FD20/10 (PRODUCT OF ALLURACLARITY FAMILY)722029
Not applicable.ALLURA XPER FD10001443
ALLURA XPER FD10
Not applicable.ALLURA XPER FD10 (PRODUCT OF ALLURACLARITY FAMILY)722026

Issue

A problem has been identified in the Philips Allura XPER and Azurion systems that could pose a risk for patients. Philips has discovered that the wireless foot switch can suddenly stop responding due to a firmware bug when a number of ambient conditions coexist, such as EMC disturbance and the presence of other wireless devices in the room. The wireless footswitch is an option to philips allura XPER and Azurion systems to release live Fluoroscopy and exposure. Recovery of this error can be done by means of a cold restart of the system [power down, power up], which restores the functionality of the base station.

Recall date: 2021-10-25