Health Products

FIVE S 3.5X65, STERILE FOR SINGLE USE

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Ear, nose and throat
Product: FIVE S 3.5X65, STERILE FOR SINGLE USE
Issue: Medical devices - Lack of efficacy
What to do: Contact the manufacturer if you require additional information
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
More than 10 numbers, contact manufacturer.FIVE S 3.5X65, STERILE FOR SINGLE USE091361-01
091361-06

Issue

For products affected, the required sterility assurance level was not achieved during recent testing of the sterilization process performed by the contract sterilizer. The testing was initiated as a result of a regular re-qualification review. The affected products therefore may not be "sterile" as claimed.

Recall start date: 2021-08-09