· Published by Health Canada · Official notice
| Lot or serial number | Affected Products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | FIVE S 3.5X65, STERILE FOR SINGLE USE | 091361-01 091361-06 |
For products affected, the required sterility assurance level was not achieved during recent testing of the sterilization process performed by the contract sterilizer. The testing was initiated as a result of a regular re-qualification review. The affected products therefore may not be "sterile" as claimed.
Recall start date: 2021-08-09