Health Products

SYNGO.PLAZA

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: SYNGO.PLAZA
Issue: Medical devices - Inaccurate test or measurement results
What to do: Contact the manufacturer if you require additional information
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
More than 10 numbers, contact manufacturer.SYNGO.PLAZA11293724

Issue

MRI images with certain values in the Dicom tags rescale slope (0028,1053) and rescale intercept (0028, 1052) can show wrong statistical values 'min/max' (maximum and minimum grayscale value in the ROI) and 'Mean/STD D' (mean value and standard deviation of the grayscale values) in the ROI, which can lead to an inconsistent, conflicting holistic view on the patient

Recall start date: 2021-09-30