Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: SYNGO IMAGING - SYSTEM IUK
Issue: Medical devices - Inaccurate test or measurement results
What to do: Contact the manufacturer if you require additional information
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
1414
1457
SYNGO IMAGING - SYSTEM IUK10 014 063

Issue

MRI images with certain values in the Dicom tags rescale slope (0028,1053) and rescale intercept (0028, 1052) can show wrong statistical values 'min/max' (maximum and minimum grayscale value in the ROI) and 'Mean/STD D' (mean value and standard deviation of the grayscale values) in the ROI, which can lead to an inconsistent, conflicting holistic view on the patient.

Recall start date: 2021-09-29