Health Products

ALOKA ARIETTA 850

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Obstetrics and gynaecology
Product: ALOKA ARIETTA 850
Issue: Medical devices - Inaccurate test or measurement results
What to do: Contact the manufacturer if you require additional information
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
S/N G3042822
S/N G3042825
S/N G3042826
ALOKA ARIETTA 850ALOKA ARIETTA 850

Issue

Two issues with the diagnostic ultrasound system Arietta 850 (software version V1.0.0 - V4.1.3) Used in combination with a Fujifilm (formally Hitachi) Transducer: (1) the calculation result of eroa is displayed incorrectly when measuring lvot flow, lvot and MR flow. This error also appears in the user manual. (2) The calculation result for MVA (VTI) is displayed incorrectly when measuring MV.

Recall start date: 2021-10-26