Health Products

ARTIS ZEE and Q

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: ARTIS ZEE and Q
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
159022
159031
159044
159048
ARTIS ZEE MULTI-PURPOSE10094139
110070
110097
ARTIS Q10848281

Issue

If SID (source-to-image distance) lift movement is activated and X-ray shall be released simultaneously then X-ray is not possible and the message "no X-ray, try again" is displayed. X-ray is not possible. However, after stopping SID lift movement and releasing the X-ray switch, X-ray is possible again.

Recall start date: 2021-11-03