SYNCHRON SYSTEMS CHOLESTEROL (CHOL) REAGENT

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices
Product: SYNCHRON SYSTEMS CHOLESTEROL (CHOL) REAGENT
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
M106450SYNCHRON SYSTEMS CHOLESTEROL (CHOL) REAGENT467825

Issue

Beckman coulter has identified that Synchron Systems Cholesterol Reagent lot M106450 may fail with out of calibration range low (OCR low) and/or back to back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. The continued use of Synchron Systems cholesterol reagent lot M106450 may lead to: -Further failed QC and/or calibrations -Erroneous low Cholesterol patient results

Recall start date: 2021-11-08