· Published by Health Canada · Official notice
| Lot or serial number | Affected Products | Model or catalog number |
|---|---|---|
| Not applicable | ALLURA XPER FD10 | 001443 ALLURA XPER FD10 |
| Not applicable | ALLURA XPER FD20/10 (PRODUCT OF ALLURACLARITY FAMILY) | 722029 |
A problem has been identified in the Philips Allura Xper and Azurion systems that could pose a risk for patients. Philips has discovered that the wireless foot switch can suddenly stop responding due to a firmware bug when a number of ambient conditions coexist, such as EMC disturbance and the presence of other wireless devices in the room. The wireless footswitch is an option to Philips Allura Xper and Azurion systems to release live Fluoroscopy and exposure. Recovery of this error can be done by means of a cold restart of the system [power down, power up], which restores the functionality of the base station.
Recall start date: 2021-10-25