· Published by Health Canada · Official notice
| Lot or serial number | Affected Products | Model or catalog number |
|---|---|---|
| 11032 11038 11042 11044 11048 11064 11079 | BIOGRAPH VISION | 10749670 10749672 |
| 91147 | BIOGRAPH MCT-S (20) | 10534160 |
| 94759 | BIOGRAPH HORIZON - 3R | 10532746 |
| 11002 11010 60036 | BIOGRAPH MCT FLOW-4R | 10529159 10529161 |
| 11013 | BIOGRAPH MCT FLOW EDGE-4R | 10528955 |
| 11058 11069 60036 | BIOGRAPH MCT-S WITH TRUE V (40) | 10248671 |
| 59002 | BIOGRAPH MCT FLOW-3R | 10528956 |
| 11030 | BIOGRAPH MCT-S WITH TRUE V (64) | 10248672 |
| 60029 | BIOGRAPH MCT-S (40) | 10248668 |
There is potential patient fall hazards could occur when the PHS pallet/foot extension is improperly assembled by the operator. PHS pallet/foot extension patient fall hazards could occur when the extension is improperly used to position the patient head-first. The pallet/foot extension should only be used for positioning the patient's feet. To avoid this issue, customers are advised to ensure proper operator assembly of the PHS pallet/foot extension as instructed in the operator's manual and as depicted on the extension label. And to only use the PHS pallet/foot extension for its intended purpose of feet first imaging as instructed in the operator's manual. The operator manual cautions of the improper use of the positioning aids and notes to only use the intended purpose. The foot extension itself is labelled and instructions of proper use of extension are also provided.
Recall start date: 2021-11-08