· Published by Health Canada · Official notice
| Lot or serial number | Affected Products | Model or catalog number |
|---|---|---|
| Not applicable. | INVACARE TDX SP2 WITH LINX GYRO | TDXSP2 TDXSP2-CG TDXSP2-MCG TDXSP2HD TDXSP2HD-CG TDXSP2HD-MCG TDXSP2V TDXSP2V-HD TDXSP2X-CG TDXSP2X-MCG |
Power wheelchairs with a linx Gyro component running firmware version 6.1.2 Can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile.
Recall start date: 2021-11-12