· Published by Health Canada · Official notice
| Lot or serial number | Affected Products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | Puritan Bennett 980 Series Ventilator | 980S1ENDICU 980S3ENDICU 980U1ENDICU |
This voluntary medical device correction is being taken following the investigation of six customer reports stating that a total of four ventilators became inoperable during use. One of these reports indicated that, following the ventilator becoming inoperable while providing mechanical ventilation, the patient experienced cardio-respiratory complication, and passed away three days later. The investigation identified the root cause for these reports to be a manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly. Only the affected serial numbers are impacted by this medical device correction.
Recall start date: 2021-11-09