· Published by Health Canada · Official notice
| Lot or serial number | Affected Products | Model or catalog number |
|---|---|---|
| S/N: 30011 S/N: 30036 | QIAcube Connect MDX | 9003070 |
The information detailed on the screen when the info icon is clicked is incorrect and should state that the volume is exact, not a minimal volume. This could cause a user to pipette an incorrect volume which could lead to mistaken results or sample loss. The main screen and user manual are correct but this supplemental information screen could confuse a user. Discovered internally at Qiagen.
Recall start date: 2021-11-18