Atellica Ch 930 Analyzer

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices
Product: Atellica Ch 930 Analyzer
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
More than 10 numbers, contact manufacturer.Atellica Ch 930 Analyzer11067000

Issue

Siemens Healthcare Diagnostics has identified issues with the Atellica CH 930 Analyzer, which are installed with Atellica solution software (SW) versions V1.25.1 and lower when using open channelassays. Siemens Healthineers assays are not impacted.

Issue 1: TDEF (test definition) parameters for open channel assays reverting to default values. If TDEF parameters have been edited for an open channel assay under the CH test definition screen, the next time an open channel configuration screen is edited, it may overwrite the previously edited open channel TDEF parameters on the CH test definition screen with default values.

Issue 2: On board stability (OBS) not updating with manual changes. When the customer manually changes the onboard stability (OBS). For reagent packs for open channel assays, the operator will only see a change for reagent 1-Well-1 with no change to R1-Well-2, R2-Well-1, and R2-Well-2. The system may continue to use the reagent past its OBS date leading to potential erroneous patient sample results.

Recall start date: Nov 24, 2021