Health Products

Syncvision System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Syncvision System
Issue: Medical devices - Inaccurate test or measurement results
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
Not applicableSyncvision System400-0100.10

Issue

If prior to the IFR/FFR pullback on the intrasight system, FFR measurement(s) were performed within the same syncvision procedural session, the co-registered distal value will be different than the distal value displayed in the philips IFR/FFR system. This error will appear "insufficient data, distal segment is not co-registered" the risk of suboptimal patient treatment due to falsely elevated data on the initial trendline for the IFR/FFR co-registration would be greatest from inexperienced interventionalists.

Recall start date: Dec 10, 2021

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime