Health Products

Acuson Sequoia Diagnostic Ultrasound System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Acuson Sequoia Diagnostic Ultrasound System
Issue: Medical devices - Inaccurate test or measurement results
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
800343Acuson Sequoia Diagnostic Ultrasound System11148775

Issue

System overestimates the circumference when the manual trace function is used. the trace circumference value is overestimated; however, the area value is accurate. the issue is present in Sequoia's VA20A/VA25 onward as the VA20A introduced a new calculation design that did not include an optimization step of discarding pixel points that are too close together, causing the close pixel points into the accumulated circumference/distance value makes the final value highly sensitive to small trackball motions. Recall start date: 2021-12-16

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime