· Published by Health Canada · Official notice
| Lot or serial number | Affected Products | Model or catalog number |
|---|---|---|
| 1069101301 1069180401 1069184501 1069008101 1069090301 | Delfia Xpress Free Hcgβ Kit | 6002-0010 6002-001C |
| 1068893801 1069077201 1069077101 | Delfia Xpress Papp-A Kit | 6003-002C 6003-0020 |
We have become aware that the measured analyte concentrations in occasional patient samples may be decreased with the affected Delfia Xpress kit lots. The issue may also cause failed results. Based on investigation the antifoam concentration of the tracers is incorrect at the end of the manufacturing series and therefore only certain tracer vials are affected by the issue. the issue may lead to occasional pipetting failures of the tracer during the assay procedure.
Recall start date: 2021-12-17