Endurant Iis Stent Graft System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Endurant Iis Stent Graft System
Issue: Medical devices - Manufacturing defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
More than 10 numbers, contact manufacturer.Endurant Iis Stent Graft SystemESBF3214C103EE, ESBF2314C103EE, ESBF2814C103EE, ESBF2514C103EE
More than 10 numbers, contact manufacturer.Endurant Ii Stent Graft SystemETCF2828C49EE, ETTF2323C70EE, ETBF2313C166EE, ETBF2513C145EE, ETBF2516C124EE, ETUF2814C102EE, ETBF2513C124EE, ETUF2514C102EE, ETBF2816C145EE, ETBF2813C124EE, ETCF2323C49EE, ETBF2516C166EE, ETBF2813C145EE, ETBF2813C166EE, ETBF2816C166EE, ETUF2314C102EE, ETBF2313C145EE, ETBF2316C166EE, ETBF2516C145EE, ETBF2820C145EE, ETCF2525C49EE, ETUF3614C102EE

Issue

Devices built with specific batches of taper tip assemblies have the potential for the taper tip to detach from the delivery system. Detachment of the taper tip during the implant procedure may lead to a secondary intervention to attempt to remove the taper tip, either by endovascular retrieval or surgical conversion. There is also potential that the taper tip is intentionally left behind in the patient, at the physician's discretion, if not easily removable. As of 09-Dec-2021, Medtronic has received five (5) complaints where the taper tip detached from the delivery system during the procedure. In three (3) cases the tip was retrieved successfully: One (1) through surgical conversion and two (2) by snaring. in the remaining two (2) cases the tip was intentionally left behind. No deaths have been reported from the five (5) complaints.

Recall start date: Dec 23, 2021