Optima Mr430s 1.5t - Main Unit

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Optima Mr430s 1.5t - Main Unit
Issue: Medical devices - Quality
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
403220MR1, 514697ONI, 905521ONIOptima Mr430s 1.5t - Main UnitOPTIMA MR430S 1.5T (5000-0002)

Issue

GE healthcare (GEHC) MR systems could potentially have a cryogen ventilation system that does not meet the venting requirements. Failure to have proper venting could present a safety issue if the cryogen gas is vented into the MR room during a magnet quench, potentially depriving the room of oxygen.

Recall start date: Jan 4, 2022