· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | Bd Veritor Plus Analyzer | 256066 |
Bd has received two (2) complaints reporting that the Bd Veritor Plus analyzer may have the potential to overheat and/or cause fire. Bd's investigation has found that the issue can occur when an unauthorized alternative ac power adapter, and not the one provided by Bd with the Bd Veritor Plus analyzer, is used to power the instrument.
Recall start date: Jan 5, 2022