| Lot or serial number | Affected Products | Model or catalog number |
|---|---|---|
| 4964226MR81, 450688MR1, 780679MRV, 780699MRV, 780679AMSTMOBL, 306664MR1 | Signa Voyager | SIGNA VOYAGER |
| 780964EXP, 604988MRI, 519824MR, 416431MR, 514765MRI, 604941MR1, 604822MRTS, 418691MR, 250782MR | Signa Explorer | 6433266-1 |
| 403541MR, 604985MR1, 450670MR2 | Optima Mr360 - Main Unit | 5433265-1, 5312366 (M50002BB) |
| 514747MRI, 604202MR, 514341MRI, 604520MR2 | Signa Hde 1.5t Mri System - Main Unit | S7505KY |
| More than 10 numbers, contact manufacturer. | Signa Artist | SIGNA ARTIST |
GE healthcare (GEHC) MR systems could potentially have a cryogen ventilation system that does not meet the venting requirements. Failure to have proper venting could present a safety issue if the cryogen gas is vented into the MR room during a magnet quench, potentially depriving the room of oxygen.
Recall start date: Jan 4, 2022