Health ProductsType I recall

Liko M220 / M230 Mobile Lift

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: Liko M220 / M230 Mobile Lift
Issue: Medical devices - Increased risk or new adverse events
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
More than 10 numbers, contact manufacturer.Liko M220 / M230 Mobile Lift2050010, 2050015

Issue

Hillrom received four (4) complaints between 29-Sep-2020 and 09-Apr-2021 of the Slingbar 450 becoming detached from the upper bracket of the Slingbar attachment on Liko M230 Mobile Lifts. The Slingbar is attached to the lifting arm through two brackets. These two brackets are attached to each other by a bolt and a sling plate that allows for rotation of the sling bar in one axis relative to the upper bracket.

The investigation identified the root cause of the failure to be that only one (1) instead of two (2) plastic bushings had been installed during production. The absence of the lower plastic bushings can lead to the center bolt wearing on the sling plate linkage mechanism. Excessive wear can result in inadequate attachment of the Slingbar to the lift, causing it to become detached and resulting in the patient being dropped from the lift.

Recall start date: Jan 13, 2022