Health Products

St80i Stress Test System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: St80i Stress Test System
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
St80i Stress Test SystemMore than 10 numbers, contact manufacturer.860343

Issue

When patient demographic information is entered into the system and the patient information is saved or exported to a folder, in very rare situations, the demographic information that is displayed is that of the previous patient instead of the current patient resulting in patient data mismatch. There have been complaints related to this issue, but to date there have been no reports of patient injury.

Recall start date: Dec 20, 2021

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime