Details

Alert type: Health product recall
Category: Health products - Medical devices - Haematology
Product: Sight Olo Products
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Sight Olo-E1up to and including v2.59.3OLO-E1
Sight Olo-U1 Hematology Analyzerup to and including v2.59.3OLO-U1

Issue

A software defect was discovered, whereby in certain circumstances, a change to a reference range value that was canceled by the operator could nevertheless be saved, which may lead to incorrect reference range being set on the device. The software version of the Olo analyzer will be updated to v2.63.

Recall start date: Jan 18, 2022