Health Products

Benchmark Ultra Stainer Module

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Pathology
Product: Benchmark Ultra Stainer Module
Issue: Medical devices - Manufacturing defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Benchmark Ultra Stainer ModuleMore than 10 numbers, contact manufacturer.N750-BMKU-FS

Issue

The manufacturer reported a fluid leak within the Benchmark Ultra instruments that created an electrical short on the instrument's general purpose input/output (GPIO) board's printed circuit board (PCB). The event was localized to the J1 connector of the GPIO board in the lower rear portion inside the instrument and there was no injury.

Recall start date: Jan 18, 2022