LiquiBandFIX8® Open Hernia Mesh Fixation Device

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: LiquiBandFIX8® Open Hernia Mesh Fixation Device
Issue: Medical devices - Premature failure
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
LiquiBandFIX8® Open Hernia Mesh Fixation DeviceP00221182, P00221313, P00221396, P00221600, P00221756, P00221983, P00225605, P00225801, P00226740FX002

Issue

During verification testing, a non-conformance was detected. Failure mode was uncontrolled leakage of glue from the distal tip of the Fix8 Open (FX002) device.

Recall start date: Feb 3, 2022