Health ProductsType I recall

Liquiband Fix8 Open Hernia Mesh Fixation Device

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: Liquiband Fix8 Open Hernia Mesh Fixation Device
Issue: Medical devices - Manufacturing defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Liquiband Fix8 Open Hernia Mesh Fixation DeviceP00221600, P00225605, P00221983, P00221396, P00221182, P00225801, P00221756, P00226740, P00221313FX002

Issue

During verification testing, a non-conformance was detected. Failure mode was uncontrolled leakage of glue from the distal tip of the Fix8 open (FX002) device.

Recall start date: Feb 3, 2022