Health Products

Vidas High Sensitive Troponin I (Tnhs)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Vidas High Sensitive Troponin I (Tnhs)
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Vidas High Sensitive Troponin I (Tnhs)All lots.415386

Issue

Since January 2021, based on complaints received from the field related to critical non-repeatable overestimated results (changing interpretation according to the cutoff value, 19 ng/l, in the frame of the three hours algorithm that lead to a change in clinical interpretation) that lead to false positive results by some customers, Biomérieux initiated an investigation that is still ongoing.

It is important to keep in mind that it is commonly acknowledged among experts that both conventional and high sensitive troponin assays can lead to abnormally non-repeatable overestimated values especially in the lower part of the measuring range.

Recall start date: Jan 31, 2022

Get told about the next health product recall.
Get health product recall alerts