Health Products

Raystation 4-11b

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Raystation 4-11b
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Raystation 4-11bversion 8.1.2, version 10.0.1, version 7.0.010B (10.1), 7, 8B, 10A (10.0)

Issue

In rare cases, the SSD (source to skin or surface distance) calculation algorithm will miss the entry point of the ROI and instead calculate the distance to the exit point, resulting in an incorrect SSD. The SSD will be incorrect in UI, Dicom export, plan report, and scripting. If the bug is triggered when an intended SSD is entered in Raystation, the intended SSD will be displayed and exported, but the actual SSD of the plan is shorter than intended.

Recall start date: Feb 11, 2022