Health Products

Artis

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Artis
Issue: Medical devices - Mechanical defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
More than 10 numbers, contact manufacturer.Artis Zee Biplane10094141
140409Artis Zee Floor Mn10094142
More than 10 numbers, contact manufacturer.Artis Zee Multi-Purpose10094139
More than 10 numbers, contact manufacturer.Artis Zee Ceiling10094137
More than 10 numbers, contact manufacturer.Artis Q10848280, 10848281, 10848282, 10848283
161003, 160846, 160423, 160354, 160936, 161020, 160109Artis Zeego Multi-Axis10280959
123214, 123048, 123506Artis Q.Zen10848353, 10848354, 10848355
More than 10 numbers, contact manufacturer.Artis Zee Floor10094135

Issue

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the x-ray tube is no longer sufficiently cooled and the system will display the message "tube hot, have a break". Several minutes later the system will block x-ray to prevent further damages and displays the message "no xray: tube too hot".

Recall start date: Feb 25, 2022