Health Products

Exalt Model D Single-Use Duodenoscope

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Gastroenterology and urology
Product: Exalt Model D Single-Use Duodenoscope
Issue: Medical devices - Revised instructions for use
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Exalt Model D Single-Use DuodenoscopeAll lots.M00542420, M00542421, M0054242CE0, M0054242CE1

Issue

Due to the higher observed occurrence of esophageal perforations compared to the rate reported in literature, a product advisory is recommended to make users aware of the events and to reiterate and supplement the warnings and directions related to perforations in the product labeling, and to communicate the intent to update the instructions for use with this information.

Recall start date: Mar 3, 2022