Health Products

Rotawire Drive Guidewire And Wireclip Torquer

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Rotawire Drive Guidewire And Wireclip Torquer
Issue: Medical devices - Quality
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Rotawire Drive Guidewire And Wireclip Torquer, Extra SupportMore than 10 numbers, contact manufacturer.H74939463005
Rotawire Drive Guidewire And Wireclip Torquer, FloppyMore than 10 numbers, contact manufacturer.H74939462005

Issue

Boston Scientific is conducting a removal of specific batches of rotawire drive and wireclip torquer after determining that a small number of pouches containing these products may have a weak seal, which could open during normal shipping and handling and thus compromise the sterile barrier. Use of a rotawire drive product from a package with a sterile barrier breach may present a risk of infection. The most serious health risk that could occur from use of a non-sterile rotawire drive is a post-procedure endocarditis or access site infection that could be managed with medical evaluation and treatment with oral or intravenous antibiotics. We have received one complaint related to this issue and there have been no patient harms or adverse events reported.

Recall start date: Mar 8, 2022