Architect "I" System - Stat Myoglobin Assay

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Architect "I" System - Stat Myoglobin Assay
Issue: Medical devices - Unauthorised device
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Architect "I" System - Stat Myoglobin Assay1660002K43-01

Issue

The purpose of this recall is to inform the customer of a product recall for architect stat myoglobin calibrators, ln 2k43-01, lot number 166000. They have identified that this calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in the instructions for use (IFU).

Recall start date: Mar 10, 2022