Affected ProductsLot or Serial NumberModel or catalog number
Immulite 2000 System-Human Chorionic Gonadotropin (Hcg) Assay466, 463L2KCG2, L2KCG6

Issue

Siemens Healthcare Diagnostics Inc. has confirmed the potential for falsely elevated hcg results due to sample carryover. This can be observed when a sample is assayed for HCG immediately after an undiluted sample with a HCG value of GT;5000 MIU/mL. This issue impacts serum and urine patient samples, as well as quality control samples and adjustors. The effect of carryover varies based on the concentration of the high HCG sample.