Prismax Control Unit

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Gastroenterology and urology
Product: Prismax Control Unit
Issue: Medical devices - Inaccurate test or measurement results
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Prismax Control UnitAll lots.955558, 955725

Issue

Baxter Corporation is issuing a correction for the Prismax. If the operator initiates therapy by entering a value other than the default value for the patient gain/loss limit or the return pressure drop limit and uses the same patient function, the Prismax system will return the patient gain/loss limit or the return pressure drop limit to the default value, rather than the non-default value entered at the initiation of treatment.

Recall start date: Mar 17, 2022