Health Products

Respironics Ventilator

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Respironics Ventilator
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Respironics V60 Plus VentilatorAll lots.1137276
V60 Ventilator, Symbols Only, With Cflex And Avaps OptionsAll lots.1053614
Respironics V680 VentilatorAll lots.850011

Issue

All V60/V60 plus and V680 units have been identified to have potential issue that could affect the main electrical circuit ("35V rail") powering the ventilator and alarm. In some cases, this issue may result in either one of the following scenarios: 1. The ventilator ceases to operate, activating both visual and audible alarms. 2. The ventilator ceases to operate and does not activate either a visual or audible alarm causing a "silent shutdown".

Recall start date: Mar 17, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime