· Published by Health Canada · Official notice
| Affected Products | Lot or Serial Number | Model or catalog number |
|---|---|---|
| Respironics V60 Plus Ventilator | All lots. | 1137276 |
| V60 Ventilator, Symbols Only, With Cflex And Avaps Options | All lots. | 1053614 |
| Respironics V680 Ventilator | All lots. | 850011 |
All V60/V60 plus and V680 units have been identified to have potential issue that could affect the main electrical circuit ("35V rail") powering the ventilator and alarm. In some cases, this issue may result in either one of the following scenarios: 1. The ventilator ceases to operate, activating both visual and audible alarms. 2. The ventilator ceases to operate and does not activate either a visual or audible alarm causing a "silent shutdown".
Recall start date: Mar 17, 2022