Health Products

Sonialvision G4

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Sonialvision G4
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Sonialvision G441C351653001, 41C3534B4001503-78000-22, 503-78000-52

Issue

Following Shimadzu's investigation, it was noted that there is a potential issue with the system software that could cause: 1) The x-ray irradiation rate (pulse rate) of pulse fluoroscopy to be different then set value, 2) The beam hardening filter during fluoroscopy to be different then set value, 3) The calculated dose displayed on the screen to be higher than actual dose (the correct does is displayed if an area dosimeter is installed). The malfunctions could lead to improper measurements of pulse rate, setting of beam hardening filter and/or inaccurate dose displayed on display screen.

Recall start date: Mar 23, 2022