Health Products

Sara Plus

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Sara Plus
Issue: Medical devices - Increased risk or new adverse events
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Sara PlusMore than 10 numbers, contact manufacturer.HEP1001-US, HEP0001-US

Issue

Since Nov, 2020, Arjo has become aware of 2 customer complaints on smoke and flames coming out of the lift. In both events the extent of flames was limited to the lift and no health consequences were reported. The product (devices and spare parts shipped between Oct 2016 - Jan 2022) in the market are recognized as potentially affected with the risk of cover ignition. The investigation has revealed that the failure is attributed to the insufficient design of Sara Plus PCB that is prone to damage during use.

Recall start date: Mar 28, 2022