Health ProductsType I recall

Medfusion Syringe Infusion Pumps

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Medfusion Syringe Infusion Pumps
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Medfusion 3500 Syringe Infusion PumpMore than 10 numbers, contact manufacturer.3500-0600-50, 3500SD-500, 3500VX-500, 3500VX-415, 3500, 3500-415, 3500-306, 3500-414, 3500-500, 3500SD, 3500VX-306
Medfusion 4000 Syringe Infusion PumpMore than 10 numbers, contact manufacturer.4000-0100-51, 4000-0101-50, 4000-0105-51, 4000-0601-01, 4000-0101-295, 4000-0101-95, 4000-0106-00, 4000-0106-01, 4000-0101-78, 4000-0100-50, 4000-0100-295, 4000-0100-95, 4000-0101-51, 4000-0100-249

Issue

Smiths Medical became aware of eight (8) issues-(identified below) that can potentially occur that may impact therapy delivery in specific Medfusion 3500 and 4000 syringe infusion pumps. 1. Primary Audible Alarm (PAA) 2. Unanticipated depleted battery alarms 3. Time Base Alarm 4. Intermittent Volume Over Time (IVOT) - Infusion continues after system fault 5. Clearing of Program Volume Delivered (PVD) 6. False Alarm for Rate below recommended minimum for syringe size 7. Incorrect bolus or loading dose time display 8. Domain Name Server (DNS) port 1001 and when some of these issues happen, It may lead to delay in therapy or interruption of therapy, which may result serious injuries to patients.

Recall start date: Mar 28, 2022