Health ProductsType I recall

Fabian Hfoi Ventilator

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Fabian Hfoi Ventilator
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Fabian Hfoi VentilatorMore than 10 numbers, contact manufacturer.113001, 7200.I

Issue

Bias flow selection for NO system usage is missing from the user interface of Fabian HFOi Ventilators following incorrect selection of device configuration during a software update. This is caused by human error and there is no defect in the device or its software.

Recall start date: Mar 28, 2022