LUMINOS DRF MAX, UROSKOP OMNIA MAX, MULTITOM RAX

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: LUMINOS DRF MAX, UROSKOP OMNIA MAX, MULTITOM RAX
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
MULTITOM RAX11109, 11017, 11106, 11107, 11111, 1110810860700
LUMINOS DRF MAXMore than 10 numbers, contact manufacturer.10762471
UROSKOP OMNIA MAXMore than 10 numbers, contact manufacturer.10762473

Issue

Siemens Healthcare will be distributing new operator's manuals to affected customers which are needed for the upcoming software upgrade to VF11.

5 cautions were added to the manual. Caution 1 (added two sentences to already existing caution): cause: Patient falls during standing 3d

Remedy: Always use the 3d ortho stand for standing 3d studies. Aid patient in climbing onto the 3d ortho stand. Make sure that the patient is stable enough to stand for the entire examination

Caution 2 cause: Operator or patient trips over the foot switch cable or the foot switch itself

Remedy: position the foot switch and the associated cable on the floor so that nobody trips.

Caution 3: cause: the detector holder is not securely fastened

Remedy: make sure the detector holder is securely fastened

Caution 4: cause: riding on wheeled accessories

Remedy: Do not ride or allow anyone else to stand, sit or ride on wheeled accessories while they are being rolled.

Caution 5: cause: unauthorized access to the system

Remedy: the hospital is responsible for network security at the site. Set up firewalls and user account password protections. Do not allow users to change configuration files. Update virus protection software as required.

Siemens Healthcare will be distributing new operator's manuals to affected customers which are needed for the upcoming software upgrade to VF11.

Caution: cause: patient cannot maintain position or becomes unconscious Remedy: check that the patient is physically able to hold the examination position, especially when standing. Choose a different position or use positioning aids for immobilizing weak or very unstable patients. Caution cause: unauthorized access to the system remedy: The hospital is responsible for network security at the site. set up firewalls.

Siemens Healthcare will be distributing new operator's manuals to affected customers which are needed for the upcoming software upgrade to VF11. This update delivers a new manual to customers to be able to use the upcoming software version VF11. A caution was added to the manual to prevent unauthorized access to the system.

Recall start date: Apr 5, 2022