· Published by Health Canada · Official notice
| Affected Products | Lot or Serial Number | Model or catalog number |
|---|---|---|
| Disposable Grasping Forceps | More than 10 numbers, contact manufacturer. | FG-51D |
Olympus has received complaints that the grasping portion of FG-51D forceps can be difficult to open and close. Olympus reviewed approximately 200 complaints globally over the last five years. Subsequent complaint investigation has confirmed that the FG-51D forceps do not comply with Olympus standards for the amount of force required to open and close the forceps.
Recall start date: Mar 31, 2022