Health Products

Karl Storz Se & Co. Kg Devices

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Ear, nose and throat, Health products - Medical devices - Gastroenterology and urology
Product: Karl Storz Se & Co. Kg Devices
Issue: Medical devices - Labelling and packaging
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
More than 10 numbers, contact manufacturer.Flexible Uretero-Reno Fiberscope11278A2, 11278AU1
More than 10 numbers, contact manufacturer.Flexible Pediatric Cystoscope11274BCU1
More than 10 numbers, contact manufacturer.Rhin0-Laryngo Fiberscope11001RD1
More than 10 numbers, contact manufacturer.Flexible Video-Uretero-Renoscope Flex-Xc11278VSU, 11278VU, 11278V/VU, 11278VSA, 11278VSUA, 11278VSE, 11278VSUE, 11278VA, 11278VUA
More than 10 numbers, contact manufacturer.Flexible Cystoscope11272VA, 11272VU, 11272C2, 11272VUE-R, 11272CU1, 11272C1, 11272V, 11272VUE, 11272VUA
More than 10 numbers, contact manufacturer.Broncho Fiberscope11005BC1
More than 10 numbers, contact manufacturer.Ureteroscope11278AC1, 11278AC2, 11278ACU1
More than 10 numbers, contact manufacturer.Flexible Digital Video Cystoscope11272VNU, 11272VN
More than 10 numbers, contact manufacturer.Flexible Video Cystoscope Hd-View11272VH-TL, 11272VHU-TL

Issue

This recall notice describes a labelling update to correct the instructions for use for the flexible endoscopes identified to advise that high-level disinfection and, for specific models, liquid chemical sterilization, should not be used as methods of reprocessing. Customers that do not have access to an appropriate sterilization method recommended in the applicable instructions for use for these endoscopes should discontinue use of the endoscopes and return them to Karl Storz as described.

Recall start date: Apr 6, 2022