Health Products

Zeus Infinity Empowered (Anesthesia Workstation)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Zeus Infinity Empowered (Anesthesia Workstation)
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Zeus Infinity Empowered (Anesthesia Workstation)More than 10 numbers, contact manufacturer.MK01703

Issue

Draeger has become aware of cases in which the diva sevoflurane modules (part number:MK01703) stopped agent delivery during application of high fresh gas flow and high agent percentage.

Recall start date: Apr 6, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime