Health Products

H7l Outbound Results Interface

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: H7l Outbound Results Interface
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
H7l Outbound Results InterfaceHL7-REO v18.4.0.17HL7 OUTBOUND RESULTS INTERFACE

Issue

One client reported that HL7 files transferred to an outbound results interface (specific to one client) would sometimes contain more than one patient identifier, which could lead to results being assigned to the wrong patient in a third party system.

Recall start date: Mar 3, 2022