Health Products

Ortho Optix™ Reader

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Haematology
Product: Ortho Optix™ Reader
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Ortho Optix™ ReaderNot applicable.6842222

Issue

Ortho Clinical Diagnostics received a complaint regarding an incorrectly displayed passing QC status for a test (see example in table below), without the site having processed the multiple QC kits configured for the test. Ortho confirmed that any test processed with multiple QC kits may encounter this issue. If one QC kit out of the multiple QC kits configured for the test has successfully passed, the software will automatically assign a passing QC status to the test even if the other QC kit configured for the test is not run or does not pass.

Recall start date: Apr 11, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime